详情介绍
氯仿气体检测管NO.137LL 检测范围0.3-4.5ppm ? 氯仿气体检测管NO.137LL 检测范围0.3-4.5ppm Performance Measuring Range 0.3 to 4.5 ppm Number of Pump Strokes 4 Correction Factor 1 Sampling Time 2 minutes per pump stroke Detecting Limit 0.2 ppm (n=4) Colour Change White → Pale purple Reaction Principle CHCl3 + I2O5 + H2S2O7 → Cl2 Cl2 + 3.3'-DIMETHYLNAPHTHIDINE ⅹ reaction product Coefficient of Variation 10% (for 0.3 to 1.5 ppm), 5% (for 1.5 to 4.5 ppm) Shelf Life 1 Year Corrections for temperature & humidity Temperature correction is necessary Store the tubes in the refrigerator to keep at 10?C (50?F) or below. Possible coexisting substances and their interferences (NOTE) Substance Concentration Interference Change colour by itself to Chlorine ℃6ppm + Pale purple Bromine ″4.5ppm No No Iodine ″4.5ppm No No 1,1,1-trichloroethane ? + Pale purple trichloroethylene ? + Pale purple Calibration gas generation Diffusion tube method TLV-TWA TLV-STEL Explosive range 10 ppm - - 仕様 目盛範囲: 0.3~4.5ppm〔基準n=4〕 測定範囲: 0.3~4.5ppm〔基準n=4〕 吸引回数〔n〕: 4〔基準〕 1回(100ml)の吸引時間: 2分 検知限度: 0.2ppm〔n=4〕 変色: 白色→淡紫色 温?湿度補正: 温度 有効期限: 1年(冷蔵庫保存) 指示精度: CV=5~10% (CV:変動係数=σ:標準偏差÷平均値×100) 干渉ガス ガス名 共存の場合の濃度と結果 単独の場合 塩素 2倍以上 +誤差 淡紫色に変色 1,1,1-トリクロロエタン ? +誤差 淡紫色に変色 トリクロロエチレン ? +誤差 淡紫色に変色 校正用ガス ガス拡散管法 許容濃度 管理濃度 TWA STEL 爆発範囲 3ppm 3ppm 10ppm -- --